Calibrator Set, DOSIMMUNE™

Medical device

  • Product Number: 227-40110
  • Component of DOSIMMUNE™ reagent kit

DOSIMMUNE™ (227-40101-55/227-40100-55) is registered as a CE/IVD device in the EEA. In other case, DOSIMMUNE™ is intendent for research use only, not for diagnostic.

Make a request
Certification
Qty

Description

DOSIMMUNE™ Calibrator Set is to quantify immunosuppressants (Cyclosporin A, Everolimus, Tacrolimus, Sirolimus) in human whole blood by LC-MS/MS method.
The set is composed by 6 calibrators with different concentration levels and 1 blank level.

The kit contains all material needed for quantifying the four immunosuppressant drugs.

Product available for RUO 227-40110-58 (Europe) / 227-40110-91 (Japan)

Technical Data

Content: 6 calibrators level (lyoph.) 6x500 µL and 1 blank (lyoph.) 500 µL
Parameter: Cyclosporin A, Everolimus, Tacrolimus, Sirolimus

Concentration:

ParametersBlankLevel 1Level 2Level 3Level 4Level 5Level 6
Cyclosporin A / 25 µg/L 100 µg/L 450 µg/L 1000 µg/L 1500 µg/L 1800 µg/L
Everolimus / 2 µg/L 5 µg/L 10 µg/L 15 µg/L 25 µg/L 35 µg/L
Tacrolimus / 2 µg/L 5 µg/L 10 µg/L 15 µg/L 25 µg/L 35 µg/L
Sirolimus / 2 µg/L 5 µg/L 10 µg/L 15 µg/L 25 µg/L 35 µg/L


Reconstitution: Remove carefully the cap of each vial and reconstitute each calibrator with exactly 500 µL of ultra-pure water. Shake during at least 20 minutes at ambient temperature until the content becomes homogeneous.

Storage: refer to the Instruction for use


Comment: the Calibrator Set DOSIMMUNE™ is delivered with a safety data sheets to use properly the product.